Analisis Dosis Radiasi Pasien Pada Pemeriksaan CT Scan Menggunakan Aplikasi Si-INTAN

  • Elshaday S.B Siregar UNUD
  • Gusti Ngurah Sutapa Udayana University
  • I Wayan Balik Sudarsana Udayana University

Abstrak

CT scan is widely used to diagnose the inside of the human body, so supervision is needed to ensure the health and safety of workers, patients and the public. One surveillance that can be done is to analyze the radiation dose of the patient on CT scan with the application of Si-INTAN. Data processing of the results of CT scan of the head, thorax and abdomen for ages 0-4 years, 5-14 years and ? 15 years using the Si-INTAN application. From the results of the data processing, the highest DRL DLP value and CTDIVOL values were obtained, for CT scan heads were 1732,8 mGycm and 31,92 mGy, the CT scan of the thorax was 2450,78 mGycm and 19,36 mGy, and for CT Abdominal Scans were 3968,85 mGycm and 19,35 mGy.

##plugins.generic.usageStats.downloads##

##plugins.generic.usageStats.noStats##

Referensi

[1] L. Sofiana, Johan A.E, I. Normahayu, Estimasi Dosis Efektif pada Pemeriksaan Multi Slice CT-Scan Kepala dan Abdomen Berdasarkan Rekomendasi ICRP 103, Jurnal Fisika FMIPA Universitas Brawijaya, Jurusan Fisika, Universitas Brawijaya, Malang, 2012.
[2] K. Miniati, G.N. Sutapa, I.W. Balik. S, Uji Kelayakan Pesawat Sinar-X Terhadap Proyeksi PA (Postero-Anterior) dan LAT (Lateral) Pada Teknik Pemeriksaan Foto Thorax, Jurnal Buletin Fisika, vol.18, no. 1, 2017, pp 27-31.
[3] Silvia, Helga, Milvita D, Prasetio H, Helfi. Y, Estimasi Nilai CTDI dan Dosis Efektif Pasien Bagian Head, Thorax dan Abdomen Hasil Pemeriksaan CT Scan Merek Philips Briliance 6, Jurnal Fisika Unand, Padang, vol. 2, no.2, 2013.
[4] I.M Hendra, N.N Rupiasih, Pemantauan Dosis Serap Radiasi Sinar-X Pada Pemeriksaan Toraks, Jurnal Buletin Fisika, vol. 21, no. 1, 2020, pp 8-13.
[5] Z. Naily, Penentuan Diagnostic Reference Level (DRL) dan Kajian Pemilihan Parameter Exposure pada CT Scan Abdomen di RSUD Dr. Soetomo, Skripsi, D4 Ahli Jenjang Radiologi, Departemen Kesehatan, Fakultas Vokasi, Universitas Airlangga, Surabaya, 2018.
[6] L. Risalatul, Penentuan Local Diagnostic Reference Level (LDRL) Pasien Pediatrik pada Pemeriksaan CT Kepala Berdasarkan Nilai Size Spesific Dose Estimate (SSDE), Skripsi, D4 Ahli Jenjang Radiologi, Departemen Kesehatan, Fakultas Vokasi, Universitas Airlangga, Surabaya, 2019.
[7] Perbangkara, Hendya, Estimasi Dosis Radiasi Pada Pemeriksaan CT Angiografi Coroner, Skripsi, Program Studi Fisika, Fakultas Matematika dan Ilmu Pengetahuan Alam, Universitas Indonesia, Depok, 2011.
[8] U. Ida, Kesehatan Otak Modal Dasar Hasilkan SDM Handal, Jurnal Profesi, vol. 8, 2012.
[9] Hoffman, Matthew MD, Human Anatomy, 2014. Available from: https://www.webmd.com, diakses 1 Mei 2019.
[10] Pearce, Evelyn. C, Anatomi dan Fisiologis Untuk Para Medis, PT. Gramedia Pustaka Utama, Jakarta, 1999.
[11] R. Suryanti. L. Savitri, W. Susanto, I. Syafitri, Iswandarini, E. Kunarsih, Rusmanto, IB Gede Putra, Manual Penggunaan Si-INTAN Ver. 2.0, P2STPFRZR BAPETEN, Jakarta, 2018.
[12] J. Valentin, ICRP, Managing Patient Dose in Multi-Detector Computed Tomography (MDCT): ICRP 102, Annals of the ICRP, vol. 77, 2007, pp. 1-79.
Diterbitkan
2020-07-06
##submission.howToCite##
SIREGAR, Elshaday S.B; SUTAPA, Gusti Ngurah; SUDARSANA, I Wayan Balik. Analisis Dosis Radiasi Pasien Pada Pemeriksaan CT Scan Menggunakan Aplikasi Si-INTAN. BULETIN FISIKA, [S.l.], v. 21, n. 2, p. 53-59, july 2020. ISSN 2580-9733. Tersedia pada: <http://103.29.196.112/index.php/buletinfisika/article/view/58880>. Tanggal Akses: 18 dec. 2025
Bagian
Articles

##plugins.generic.recommendByAuthor.heading##

1 2 > >>